Can a Shaker Use Maximum Load and Maximum Speed Together?
The two maxima alone do not establish a permissible combination.
View resource →
Series: Universal Shakers
Model: SM 30 AT control
Combines a heating plate, touchscreen and programmable control with orbital motion for simultaneous heating and mixing of loads up to 25 kg.
Application / measured object:
Large-volume, heavy-load or high-viscosity samples requiring mixing under controlled heating.
SM 30 AT control combines orbital shaking with a heating plate for samples requiring simultaneous heating and mixing; its maximum load is 25 kg.
| Parameter | SM 30 AT control |
|---|---|
| Motion | orbital |
| Shaking platform | 560 x 400 mm |
| Max. load | 25 kg |
| Shaking speed | 15 – 300 rpm, in steps of 5 |
| Stroke | 26 mm |
| Runtime | programmable / continuous |
| Temperature range | up to 120°C |
| Control accuracy | approx. +/- 1°C |
| Order No. | 6200 000 |
| Electrical supply | 230 V, 50 / 60 Hz 115 V version on request |
| Enclosure protection | IP 21 |
| Max. heating power | 400 W |
| Ambient temperature | 5°C to 50°C |
| Relative humidity | ~ 85 % |
| Dimensions (w x d x h) | 680 x 610 x 160 mm |
| Weight | 40 kg |
Load limits vary with speed, motion, stroke and rack tiers. Consult the original loading guide in the resource section below for the full combinations.
Basic equipment includes the heated shaking platform and touch display, but not the rack system.
| Optional accessory | Order number |
|---|---|
| Rack System Combifix SM A | 0050 154 |
| Rack System Combifix SM B | 0050 155 |
| Rack System Combifix SM C | 0050 156 |

Large-volume, heavy-load or high-viscosity samples requiring mixing under controlled heating.
Select configuration by vessel temperature resistance, plate temperature, total load and fixation; the ordinary SM 30’s 30 kg load does not apply.
Product selection and application support
Tell us about your sample, application and target parameters. The Labsun team will help you select a suitable instrument or solution.