Process Refractometer Selection: Confirm the Operating Conditions

2026-10-02English

A process refractometer participates directly in process measurement, so a concentration range alone is insufficient. The A.KRÜSS Optronic requirements form separates inline/bypass installation, temperature, pressure, medium and connections, providing a basis for an engineering checklist.[1]

Confirm the point and installation arrangement

Inline and bypass measurement have different connection and operating requirements. The form specifies different pressure conditions for PR21S and PRB21S; shared refractive-index measurement does not make them interchangeable at the same connection and pressure.

Form example Listed condition Meaning
PR21S inline Gauge pressure below 9 bar Check the actual pressure for that design
PRB21S bypass Gauge pressure below 2 bar Do not apply the inline limit to bypass
Continuous operating temperature 10–60 °C Different from cleaning temperature
CIP condition Listed up to 120 °C Not permission for continuous measurement at 120 °C

These conditions are from the existing Version 1.7 form for the specified arrangements, not universal process-refractometer specifications.[1]

Overall coverage versus an individual window

PR21S information covers an overall nD span of 1.32–1.52, but an individual window is 0.1000 wide, for example 1.32–1.42 or 1.37–1.47. Quoting only the total span implies every configuration covers it. Provide the expected sample interval and confirm the chosen window.[1]

PRB21S has its own range. Do not combine the two as shared series specifications. For concentration output, establish the sample system and conversion basis rather than applying one sugar relation to every medium.

Check media, materials and cleaning separately

Provide composition, state, expected temperature/pressure and cleaning media/conditions, then check the listed wetted materials. One compatible material does not establish suitability of every seal and connection.

The form provides no explosion protection. Do not add that capability or infer hazardous-duty suitability from the product name; confirm equipment and site requirements separately.[1]

Apply the checklist to the project

Record the measurement point, connections, continuous conditions, short-term cleaning conditions, window, outputs and maintenance. Recheck configuration after conditions change rather than updating one range value.

This checklist reduces selection omissions; it is not installation instructions or a guarantee of on-site measurement performance.