NEXOPART EML 200 Pure
NEXOPART

NEXOPART EML 200 Pure Sieve Shaker

Series: EML

Model: EML 200 Pure

Base-model electromagnetic sieve shaker for samples up to approximately 3 kg with two preset amplitudes.

Application / measured object:

Standard sieve analysis of powders and granulates up to approximately 3 kg.

Application area:

Product overview

The NEXOPART EML 200 Pure is the base model for standard laboratory sieve analysis, handling approximately 3 kg with 50–203 mm sieves and two preset amplitudes, fine and coarse.

Sieving principle

The EML electromagnetic drive produces controlled three-dimensional motion, accelerating material vertically from the mesh while distributing it horizontally for efficient, reproducible sieve analysis.

Key features

  • Two preset amplitudes: fine (~0.9 mm) and coarse (~2.2 mm)
  • Fixed 10-second interval and digital timer
  • Classic “eco” clamping system
  • Inspection-glass cover and cast-aluminium housing

Technical specifications

Technical specificationEML 200 Pure
Method of analysisSieving
Measurement range20 µm–125 mm
Maximum sieving material batchApprox. 3 kg
Sieve tower weightMax. 8.7 kg
Sieve diameter50–203 mm
Dry sievingYes
Wet sievingNo
Voltage110–230 V, 50–60 Hz
AmplitudeFine (~0.9 mm) and coarse (~2.2 mm)
Amplitude adjustmentControlled and self-adjusting amplitude
Timer0–99 minutes / constant operation
Operation typeInterval
IntervalFixed, 10 seconds
Program memory locationsNo
Calibratable test equipmentCalibratable; suitable for ISO 9001 quality systems
InterfaceNo
Dimensions (W × D × H)385 × 295 × 895 mm
Net weight37 kg
ColourRAL 9010, pure white
Order No.205322921 (Classic “eco”)

Sieve loading and adapter requirements

Sieve diameter (mm)Nominal height (mm)Adapter / reduction partMaximum sieve count
502520788512719
76.22520532367619
1004520532357710
1204021121434011
150502053235919
20032—15
20050—9
20332—15
20350—9

50, 76.2, 100, 120 and 150 mm sieves require the corresponding adapter kit. A dash in the 200/203 mm rows means no kit is required in this table. The model-specific sieve-tower weight limit still applies.

Applications and selection

Suitable for particle-size distribution and sieve analysis of powders and granulates in chemicals, pharmaceuticals, food, minerals, construction materials, seeds and related fields. Selection should confirm sample weight, range, sieve diameter, dry or wet operation, clamping, voltage and documentation. Labsun is the exclusive distributor for NEXOPART products in China.