Lovibond Model F Visual Colorimeter — product documents
Official English PDF documents for the listed product models, covering features, technical specifications and configuration information.
View resource →
Series: Model F
Model: Model F
Lovibond Model F uses RYBN glass standards and adjacent-field visual comparison, part number 180000, with path lengths up to 153 mm.
Application / measured object:
Visual RYBN colour measurement
The Lovibond® Model F is a benchtop visual colorimeter using calibrated Lovibond RYBN glass colour standards. Under controlled illumination approximating daylight, the viewing tube presents adjacent sample and comparison fields. Red, yellow, blue and neutral filter racks are moved until a visual match is obtained and the result is expressed in Lovibond units.
| Part number | Model | Description |
|---|---|---|
| 180000 | Model F | Standard configuration for visual Lovibond RYBN measurement. |
| Measurement method | Visual comparison through two adjacent fields |
|---|---|
| Colour scale | Lovibond RYBN |
| Measurement range | Red 0.1–79.9; Yellow 0.1–79.9; Blue 0.1–49.9; Neutral 0.1–3.9 |
| Resolution | 0.1 Lovibond unit |
| Filter racks | 3 red, 3 yellow, 3 blue and 2 neutral; 11 total |
| Light source | Pair of 12 V / 10 W tungsten-halogen lamps |
| Optical system | Prismatic system with integrated blue filter |
| Illumination | Approximates to daylight |
| Path length | 0.1–153 mm (0.004–6 in) |
| Colour type | Transparent, translucent and, with the appropriate accessory, opaque samples |
| Applicable standards | AOCS Cc 13e-92, AOCS Cc 13j-97, BS 684, IS 548 (Part 1)-2010, ISO 15305, ISO 27608, MS 252 Part 16, MS 817 Part 12 |
| Housing | Fabricated sheet steel with a tough textured paint finish |
| Power | 220 Vac / 12 Vdc power adapter |
| Dimensions and weight | 330 × 410 × 230 mm; 8.3 kg |
| CE marking | CE; confirm the scope in the declaration of conformity for the model |
Model F instrument (including 11 Lovibond RYBN glass-standard racks), spare white reference, one pair of spare tungsten-halogen lamps, one 1-inch cell and one 5.25-inch cell.

Model F is used for Lovibond RYBN colour measurement of edible oils and fats, petroleum products and waxes, beer, malts and caramels, food and beverage products and liquid chemicals. For opaque samples or a specific standard method, confirm the required accessories and configuration during selection.
Product selection and application support
Tell us about your sample, application and target parameters. The Labsun team will help you select a suitable instrument or solution.