A.KRÜSS Optronic Brochure Polarimeter
A.KRÜSS Optronic Brochure Polarimeter, English PDF document.
View resource →
Series: Polarimeter
Model: P3000
The economical P3000 completes standard optical-rotation and international sugar-scale measurements in 1–2 seconds with temperature compensation.
Application / measured object:
Standard optical-rotation and international sugar scale °Z measurement in transparent liquids.
P3000 is intended for standard polarimetry where ±0.01° accuracy is sufficient. It displays optical rotation or the ICUMSA international sugar scale within 1–2 seconds and can convert values to 20 °C.

| Item | Specification |
|---|---|
| Scales | Optical rotation (°) and international sugar scale (°Z) |
| Measurement range | ±90°; ±259 °Z |
| Measurement accuracy | ±0.01°; ±0.01 °Z |
| Measurement period | 1–2 s |
| Resolution | 0.01°; 0.01 °Z |
| Methods and users | Not available |
| Adjustment | Automatic, menu-driven |
| Parts in contact with sample | Borosilicate glass, PTFE and stainless steel 1.4305 |
| Sample transmittance | >0.1% (OD 3) |
| Wavelength | 589 nm |
| Maximum tube length | 220 mm |
| Additional feature | Temperature compensation |
| Temperature control | Retrofittable using PT31/PT80, P8020 sample-chamber bushing and temperature-controlled tube |
| Temperature-measurement accuracy/resolution | ±0.1 °C / 0.1 °C |
| Operation | 3.5-inch capacitive touchscreen, 320 × 240 pixels; keypad |
| Interface | 1 × RS-232 |
| Dimensions (W × H × D) | 645 × 200 × 360 mm |
| Weight | 28 kg |
| Operating voltage/frequency | 110–250 VAC, 50/60 Hz |
| Power consumption in measuring mode | 30 W |
| Maximum power consumption | 40 W |
For standard polarimetry and sugar analysis in food, sugar, pharmaceutical, chemical and teaching laboratories.
The official page lists −17°, +17° and +34° premium quartz control plates (±0.001° accuracy, factory certificate, PTB-certifiable) and ±17°/±34° economy plates (±0.005° accuracy, factory certificate); each must be purchased separately according to calibration requirements.
| Optional model | Tube construction | Length | Filling volume |
|---|---|---|---|
| PRG-100-E | Glass, filling funnel | 100 mm | 6 ml |
| PRG-200-E | Glass, filling funnel | 200 mm | 12 ml |
Sample-temperature sensors, CBM910 matrix printer, cable, paper and ribbon, and tube glass cover plates and rubber rings are available separately.
For standard tasks without complex method and user management. Confirm accuracy, sample transmittance, tube length, temperature compensation, external temperature control and printing.
Note: measurement period is specified after sample temperature equalisation and at OD < 1. Choose control equipment and tube material for the sample and method.
Product selection and application support
Tell us about your sample, application and target parameters. The Labsun team will help you select a suitable instrument or solution.