KRÜSS FP8000 Series Flame Photometers
A.KRÜSS Optronic

KRÜSS FP8000 Series Flame Photometers

Series: FP8000

Model: FP8400 / FP8500 / FP8600 / FP8700

The FP8000 series offers four flame-emission configurations for Na, K, Li and Ca, from manual to continuous and automatic sampling.

Application / measured object:

Flame-emission quantification of Na, K, Li and Ca in liquid samples.

Application area:

Product overview

The FP8000 series uses flame emission for rapid quantification of alkali and alkaline-earth metals, with four configurations matched to throughput and automation.

Models and core specifications

Model Positioning Sampling/automation Typical sample volume
FP8400 Economical high-precision laboratory model Manual measurement 2.5 ml
FP8500 Continuous automatic measurement 24-hour process-monitoring operation 2.5 ml
FP8600 High-throughput autosampler model Automatic supply, undiluted measurement 2.5 ml
FP8700 External automation system Sampler + diluter + mixer Microlitre range, dilution-dependent

Advantages

Element analysis and measuring principle

  • Flame atomic emission spectrometry (AES) determines Na, K, Li and Ca in aqueous solutions in one measurement step.
  • The nebulised sample enters the flame; characteristic element-emission intensity is used for concentration analysis.
  • Propane is recommended for alkali elements, with butane as an alternative; flame-photometry acetylene is recommended for alkaline-earth elements.

Models and sampling automation

  • FP8400 provides manual measurement; FP8500 supports continuous automatic measurement and process monitoring.
  • FP8600 uses an autosampler for undiluted samples; FP8700 combines sampling, dilution and mixing for diluted microsamples.
  • Up to 120 measurements per hour; modular construction supports functional expansion with retrofit kits.
FP8000-series automated sample-supply configuration.
FP8000-series automated sample-supply configuration.

Calibration and result quality

  • Linear calibration with two standards, or non-linear polynomial fitting with six to eight standards.
  • Optional 5 mmol/l lithium internal reference; caesium reference on request.
  • Accuracy is 1% at 40 ppm; precision varies by model. See the element-specific ranges and detection-limit table.

Operation and data connections

  • 8.4-inch TFT touchscreen, 800 × 600 pixels.
  • Two USB ports, Ethernet and printer RS-232; FP8500 can be upgraded with a passive 4–20 mA interface.

Technical data

Item Specification
Elements Na, K, Li, Ca
Calibration Two-point linear; six- to eight-point non-linear polynomial fitting
Drift 1% in 60 min
Optional internal reference Lithium 5 mmol/l; caesium on request
Sample volume FP8400/FP8500/FP8600: 2.5 ml; FP8700: microlitre range depending on dilution and sample tube
Combustion gas Propane, butane or flame-photometry acetylene according to element
Display 8.4-inch TFT touchscreen; 800 × 600 pixels
Interfaces 2 × USB;1 × Ethernet;1 × RS-232
FP8500 optional output Passive 4–20 mA current interface
Maximum throughput 120 measurements/hour

Element ranges and detection limits

Element FP8400/8500/8600 range FP8700 range FP8400/8500/8600 detection limit FP8700 detection limit
Na 0.01–4500 ppm 0.1–45000 ppm 0.01 ppm 0.1 ppm
K 0.01–4500 ppm 0.1–45000 ppm 0.01 ppm 0.1 ppm
Li 0.01–4500 ppm 0.1–45000 ppm 0.01 ppm 0.1 ppm
Ca 0.50–4500 ppm 5.0–45000 ppm 0.03 ppm 0.3 ppm
At 40 ppm FP8400 FP8500 FP8600 FP8700
Na/K/Li/Ca precision 0.2% 0.2% 0.4% 0.6%
Na/K/Li/Ca accuracy 1% 1% 1% 1%

Detection limits differ by element and model; select using the table above.

Application areas

For alkali and alkaline-earth element analysis in food and beverage, clinical and life science, agriculture, chemical, water and process monitoring.

Calibration standards

Lithium-reference concentrate, 140/5 solution and certified standards 1–5 are available separately. Standards 1–5 contain Na, K, Li and Ca at 10, 20, 30, 40 and 50 mg/l respectively; select and purchase for the calibration requirement.

Related accessories and consumables

Optional items include matrix printer, pressure-reducing station with filters, nebuliser and seals, nebuliser chamber, gas-mixing chamber, glass cylinders, burner insert and model-specific sample cannulas. Consumables include hose sets, cleaners and 10 ml sample tubes; a 2.5 ml diluter syringe is available for FP8700.

Service support

  • Experienced specialists provide product selection, application and operation support.
  • Calibration, maintenance and maintenance contracts can be discussed as required.

Selection guidance

First define elements, concentration range and fuel-gas conditions, then select manual, continuous-monitoring, autosampler or external-automation configuration by sample throughput.