KRÜSS DS7000-M Manual-Fill Digital Density Meters
A.KRÜSS Optronic

KRÜSS DS7000-M Manual-Fill Digital Density Meters

Series: DS7000

Model: DS7700-M / DS7800-M

DS7700-M/DS7800-M use manual syringe filling for density and concentration measurements of low- to high-viscosity samples.

Application / measured object:

Density, relative-density and concentration measurements for suitable liquids and emulsions; sample identification, quality and purity checks.

Application area:

Product overview

For manual-fill models, approximately 0.9 ml sample is injected into the oscillating U-tube by syringe, suiting laboratories with fewer samples or a priority on flexible sample changes.

Models and core specifications

Model Sample supply Density accuracy Resolution
DS7700-M Manual syringe ±0.001 g/cm³ 0.001 g/cm³
DS7800-M Manual syringe ±0.0001 g/cm³ 0.0001 g/cm³

Advantages

Sample supply, cleaning and drying

  • Manual Luer-syringe filling for low- to high-viscosity samples; inspect the viewing window for bubbles.
  • Inject suitable solvent until residue dissolves and is removed; the attached drying unit removes remaining liquid.
Manual syringe sample-supply configuration.
Manual syringe sample-supply configuration.

Oscillating U-tube and measuring scales

  • Oscillating U-tube method with borosilicate-glass and PTFE sample-contact parts; minimum measuring-cell sample volume 0.9 ml.
  • Pre-installed density, relative density, Brix, alcohol volume fraction and sulphuric-acid mass-fraction scales; unlimited user-defined scales.
  • DS7700 and DS7800 density accuracy / resolution grades are 0.001 and 0.0001 g/cm³ respectively.

Sample temperature control and adjustment

  • Integrated Peltier control: 10–40 °C, control accuracy ±0.01 °C; temperature measurement accuracy ±0.02 °C and resolution 0.01 °C.
  • Automatic menu-guided adjustment with an air-pressure sensor and temperature-controlled calibration media; measured-value stabilisation and pressure display.
  • The three-second period applies after temperature equalisation, not to the entire filling, cleaning and drying cycle.

Operation and data integrity

  • 7.0-inch, 800 × 480 capacitive touchscreen with six interface languages; keyboard, mouse and barcode-scanner support.
  • More than 10,000 methods, user management and audit trail.
  • USB, RS-232 and Ethernet with LIMS integration; data-management functions support 21 CFR Part 11 requirements.

Technical data

Item Specification
Measurement range 0–3 g/cm³
Temperature-control range 10–40 °C
Temperature-control accuracy ±0.01 °C
Measurement period 3 s after temperature equalisation
Minimum measuring-cell sample volume 0.9 ml
Pressure range (relative) 0–9 bar
Temperature-measurement accuracy ±0.02 °C
Temperature-measurement resolution 0.01 °C
Scales Density, relative density, Brix, alcohol (vol.%), sulphuric acid (mass%), unlimited user-defined scales
Method capacity More than 10,000
Adjustment Automatic, menu-guided; air-pressure sensor and temperature-controlled calibration media
Sample-contact materials Borosilicate glass, PTFE
Additional functions Audit trail, user management, measured-value stabilisation, air-pressure display
Operation and display 7.0-inch capacitive touchscreen, 800 × 480 pixels; keyboard, mouse, barcode scanner
Physical interfaces 1 × USB;1 × RS-232;1 × Ethernet
System integration LIMS
Dimensions (W × H × D) 220 × 220 × 430 mm
Instrument weight 5.3 kg
Operating voltage / frequency 100–240 V; 47–63 Hz
Measuring / maximum power 25 W / 120 W

Density accuracy is specified at 20 °C, 1013 hPa and 50% relative humidity. 0.9 ml is the measuring-cell minimum; total sample requirements also depend on tubing and sample-supply configuration.

Application areas

For low-throughput density, relative-density and concentration measurement in pharmaceutical, chemical, petrochemical, food and beverage laboratories.

Calibration standards

Certified isooctane, n-nonane, ultrapure-water and dichlorotoluene density standards are available separately. Use the temperature and reference density stated on the relevant certificate.

Related accessories and consumables

Pump and autosampler options can be chosen for the workflow. AS80 provides 18 or 36 positions; AS90 provides 53 or 89. Replacement consumables include TPE pump tubing, PTFE connectors and adaptors. Protective cover, CBM910 printer and print consumables are available separately.

Service support

  • Experienced specialists provide product selection, application and operation support.
  • Calibration, maintenance and maintenance contracts can be discussed as required.

Selection guidance

The models differ mainly in density accuracy and resolution. Choose manual filling for low sample counts and flexible changes, then match accuracy to quality-system requirements.