KRÜSS DR Series Digital Handheld Refractometers
A.KRÜSS Optronic

KRÜSS DR Series Digital Handheld Refractometers

Series: Digital handheld refractometers

Model: DR101-60 / DR201-95 / DR301-95

DR101-60, DR201-95 and DR301-95 digital handheld refractometers provide fast refractive-index and %Brix measurements in the field or laboratory.

Application / measured object:

Fast mobile measurement of refractive index nD, sucrose concentration %Brix and model-dependent scales in small liquid samples.

Application area:

Product overview

DR101-60, DR201-95 and DR301-95 are digital handheld refractometers for rapid field, production and laboratory analysis. Once a sample is applied, a result is displayed in approximately one second, eliminating subjective eyepiece interpretation.

The models differ not only in range and accuracy: DR301-95 adds a salt-content scale and two further scales on request, while protection class, dimensions, weight and power supply also differ.

Models and core specifications

Model Scales Measurement range Measurement accuracy IP code
DR101-60 Refractive index nD; sucrose concentration %Brix nD 1.3330–1.4419; 0–60 %Brix nD ±0.0005; ±0.35 %Brix IP65
DR201-95 Refractive index nD; sucrose concentration %Brix nD 1.3330–1.5318; 0–95 %Brix nD ±0.0003; ±0.2 %Brix IP64
DR301-95 Refractive index nD; sucrose concentration %Brix; salt content; two further scales on request nD 1.3330–1.5318; 0–95 %Brix nD ±0.00015; ±0.1 %Brix IP50

Advantages

Digital measurement and fast operation

  • Digital acquisition provides results independent of visual interpretation.
  • Approximately one second after temperature equalisation with only 0.3 ml sample.
  • The fully chromed measuring trough supports fast filling and cleaning.
KRÜSS digital handheld refractometer sample application
Digital handheld refractometers support rapid measurements with small sample volumes.

Automatic temperature compensation

  • DR101-60 and DR201-95 provide 10–40 °C compensation for the %Brix scale.
  • DR301-95 temperature compensation is optionally activatable and can be used with its configured scales.
  • Integrated temperature sensor with ±0.5 °C accuracy and 0.1 °C resolution.

Model positioning and scales

  • DR101-60: refractive index nD and sucrose %Brix; IP65 and cleanable under running water.
  • DR201-95: refractive index nD and sucrose %Brix with range extended to 0–95 %Brix.
  • DR301-95: refractive index nD, sucrose %Brix, salt content and two further scales on request, up to five in total.

Scale and function selection

DR301-95 extended functions

  • Three standard scales: refractive index, sucrose concentration and salt content. Two further scales, such as Oechsle (°Oe) or Baumé (°Bé), can be added on request, for up to five.
  • Configurable alarm limits; use the supplied mains adaptor for benchtop work or a 9 V battery for mobile use.
  • DR101-60 and DR201-95 %Brix compensation cannot be deactivated; DR301-95 compensation is selectable in the menu.

Technical data

Specification DR101-60 DR201-95 DR301-95
Resolution nD 0.0001; 0.1 %Brix nD 0.0001; 0.1 %Brix nD 0.0001; 0.1 %Brix
Measurement period Approx. 1 s after temperature equalisation Approx. 1 s after temperature equalisation Approx. 1 s after temperature equalisation
Minimum sample volume 0.3 ml 0.3 ml 0.3 ml
Temperature compensation 10–40 °C; %Brix scale only 10–40 °C; %Brix scale only 10–40 °C; optionally activatable
Temperature accuracy/resolution ±0.5 °C / 0.1 °C ±0.5 °C / 0.1 °C ±0.5 °C / 0.1 °C
Adjustment/operation Membrane keypad Membrane keypad Membrane keypad
Parts in contact with sample Optical flint glass SF4; fully chromed surface Optical flint glass SF4; fully chromed surface Optical flint glass SF4; fully chromed surface
Dimensions (W × H × D) 110 × 62 × 32 mm 130 × 80 × 40 mm 180 × 100 × 60 mm
Weight 160 g 180 g 500 g
Operating voltage 1.5 V (1 × AAA battery) 1.5 V (1 × AAA battery) 9 V (1 × 9 V battery; mains-adapter operation also possible)

Official notes: accuracy is specified at λ = 589 nm, 20 °C, 1013 hPa and 50% relative humidity; measurement period applies after temperature equalisation.

Display and power supply

All three models display the measured value and temperature on an LCD. DR301-95 also displays battery charge, temperature compensation and limits. DR101-60 and DR201-95 have no USB interface and do not support mains-adaptor operation; DR301-95 can use the supplied mains adaptor.

Application areas

For rapid refractive-index, Brix, salinity and concentration checks in food and beverage, pharmaceuticals, automotive and mechanical engineering, production quality control, incoming/outgoing inspection, research and teaching.

Calibration standards

Adjustment with demineralised water is supported. Refractive-index standards at nD 1.3400, 1.3900 or 1.4300 (25 °C) can be selected according to range for periodic calibration checks. Purchase standards separately as required; KALH traceable factory calibration is available.

Service support

  • Experienced specialists provide product selection, application and operation support.
  • Calibration, maintenance and maintenance contracts can be discussed as required.

Selection guidance

DR101-60 suits lightweight, waterproof basic work to 60 %Brix; DR201-95 is the general-purpose choice for 0–95 %Brix; DR301-95 suits higher accuracy, salt content or other customised scales. Also compare protection class, dimensions, weight and power supply.