A.KRÜSS Optronic Brochure Refractometer
A.KRÜSS Optronic Brochure Refractometer, English PDF document.
View resource →
Series: Abbe refractometer
Model: AR4
The AR4 is a classic manual Abbe refractometer using transmitted light, eyepiece reading and digital temperature display.
Application / measured object:
Refractive index nD and sucrose concentration %Brix measurement in transparent liquids, solutions, fats, oils and transparent solids.
The AR4 is a classic analogue Abbe refractometer using transmitted light to determine refractive index and %Brix. Scales are adjusted with a drive knob and read through an eyepiece; an integrated digital thermometer displays temperature. A PT31 circulation thermostat can be connected when temperature control is required.

| Item | Specification |
|---|---|
| Measurement range | nD 1.3000–1.7200; 0–95 %Brix |
| Measurement accuracy | nD ±0.0002; ±0.1 %Brix |
| Resolution | nD 0.0005; 0.25 %Brix |
| Wavelength | 589 nm |
| Temperature-measurement accuracy | ±0.5 °C |
| Temperature-measurement resolution | 0.1 °C |
| Temperature control | Retrofittable with a PT31 circulation thermostat |
| Scales | Refractive index nD and sucrose concentration %Brix |
| Adjustment | By adjustment screw |
| Parts in contact with the sample | Optical flint glass SF4 |
| Operation | Drive knob and eyepiece reading |
| Dimensions (W × H × D) | 100 × 270 × 190 mm |
| Weight | 2.5 kg |
| Operating voltage | 7.5 V |
| Maximum power consumption | 15 W |
Measurement accuracy is specified at λ = 589 nm, 20 °C, 1013 hPa and 50% relative humidity; temperature control requires a PT31 circulation thermostat.
Clean the prism promptly after measurement to prevent dried or hardened residue. Remove the sample with a soft, lint-free cloth, then wipe with a slightly damp cloth. After markedly different samples, check cleaning with water or an appropriate reference.
For identity, purity and concentration testing of raw materials, semi-finished and finished products, sugar-concentration measurement, incoming and outgoing goods inspection, hospital laboratories, education and pharmacies.
The AR-series nD 1.5167 calibration standard and 5 ml cinnamon oil used as contact liquid must be purchased separately according to calibration requirements.
AR4 is intended for routine or teaching measurements with manual eyepiece reading. Selection should consider sample transparency, range, reading resolution and whether PT31 external temperature control is required.