Series: KF 54 V stainless-steel cold traps

Model: KF54V-K16-Z / KF54V-K25-Z

KF54V complete sets use two-piece 1.4301 stainless-steel coldfingers with glass or stainless-steel Dewars.

Application / measured object:

Condensation of moisture, solvents or gases and vacuum-pump protection in laboratory, medical, biotechnology and vacuum applications.

Application area:

Product overview

KF54V complete sets use two-piece 1.4301 stainless-steel coldfingers with glass or stainless-steel Dewars. The KF NW50 body joint allows separate removal of the condensate chamber; vacuum-line connections are KF NW16 or KF NW25.

Key features

  • Standard stainless steel 1.4301; 1.4404 is available on request.
  • The two body sections connect through KF NW50 with a quick-release fastener for emptying and cleaning.
  • In fixed installations, the lower condensate chamber can be removed without dismantling the upper line.
  • Coldfinger pressure range extends to 10⁻⁶ mbar vacuum and up to 3 bar gauge pressure; this does not make the Dewar coolant space pressure-rated.
  • 18 C glass and DSS 2000 stainless-steel Dewars provide different coolant volumes: 1.6 L and 1.2 L respectively.
  • Additional 360/560 variants hold 300/450 ml of condensate; sets including a CO₂ basket have separate order numbers.

Construction and model illustrations

Example of KF54V with a DSS 2000 stainless-steel Dewar; select other Dewar materials from the model tables.
Example of KF54V with a DSS 2000 stainless-steel Dewar; select other Dewar materials from the model tables.
KF54V assembly schematic showing coldfinger, support ring and Dewar separately; not a scale drawing.
KF54V assembly schematic showing coldfinger, support ring and Dewar separately; not a scale drawing.
KF54V assembly schematic with a CO₂ basket; complete sets with a basket have separate order numbers.
KF54V assembly schematic with a CO₂ basket; complete sets with a basket have separate order numbers.

Technical specifications and models

Standard complete cold traps

Type Condensate volume Coolant volume Dewar type Cold traps Connections Order No.
Type KF 54V-K16-Z-18C 200 ml 1.6 liter 18 C KF NW 16 17110
Type KF 54V-K16-Z-DSS2000 200 ml 1.2 liter DSS 2000 KF NW 16 17111
Type KF 54V-K25-Z-18C 200 ml 1.6 liter 18 C KF NW 25 17112
Type KF 54V-K25-Z-DSS2000 200 ml 1.2 liter DSS 2000 KF NW 25 17113
Type KF54V-K16-Z-360 300 ml 5 liters Type S22CAL shortened KF NW 16 17108
Type KF54V-K16-Z-560 450 ml 7.5 liters Type S22CAL KF NW 16 17109

Complete cold traps with CO2 wire basket

Type Condensate volume Coolant volume Dewar type Cold traps Connections Order No.
Type KF 54V-K16-Z-18C-CO2 200 ml 1.6 liter 18 C KF NW 16 17130
Type KF 54V-K16-Z-DSS2000-CO2 200 ml 1.2 liter DSS 2000 KF NW 16 17131
Type KF 54V-K25-Z-18C-CO2 200 ml 1.6 liter 18 C KF NW 25 17132
Type KF 54V-K25-Z-DSS2000-CO2 200 ml 1.2 liter DSS 2000 KF NW 25 17133

Optional accessories and spare parts

Spare parts for sets without CO₂ basket

Description / type Order number
Cold trap S 54V-K16-Z 17115
Cold trap S 54V-K25-Z 17114
Dewar made of glass type 18 C 10220
Support ring for type 18 C 17116
Stainless steel dewar type DSS2000 2103
Support ring for DSS 2000 17117
Dewar made of glass type S22CAL 10324
Dewar made of glass type S22CAL shortened 10324-4
Support ring for type 22 CAL 1733-208

Spare parts for sets with CO₂ basket

Description / type Order number
Cold trap S 54V-K16-Z 17115
Cold trap S 54V-K25-Z 17114
Dewar made of glass type 18 C 10220
Support ring for type 18 C with CO2 grid 17120
Stainless steel dewar type DSS2000 2103
Support ring for DSS 2000 with CO2 grille 17121

Applications

Condensation of moisture, solvents or gases and vacuum-pump protection in laboratory, medical, biotechnology and vacuum applications.

Selection guidance

Check external flange, condensate volume, Dewar material and coolant volume, then whether a CO₂ basket is needed. Distinguish the KF NW50 body joint from the KF NW16/NW25 line connections.