Herolab HiCen GT Large-Capacity Benchtop Centrifuge
Herolab

Herolab HiCen GT Large-Capacity Benchtop Centrifuge

Series: Red Line

Model: HiCen GT

Non-refrigerated large-capacity benchtop centrifuge supporting up to 4×1000 ml with swing-out or angle rotors.

Application / measured object:

For large-capacity clarification, cell harvesting, pelleting and production quality control at ambient temperature.

Application area:

Product overview

HiCen GT is a non-refrigerated large-capacity benchtop centrifuge supporting up to 4×1000 ml with either swing-out or angle rotors.

Models and performance

Part number Model Max. swing-out capacity Max. angle-rotor capacity Max. speed Max. RCF Temperature range
21 11 155 HiCen GT 4×1000 ml 4×1000 ml 14,000 rpm 21,910 ×g Non-refrigerated

Capacities, speeds and RCF values are maxima available with compatible rotors; the actual configuration depends on the selected rotor, adaptor and tube format.

Technical specifications

Swing-out rotors

Part number Rotor model Maximum capacity HiCen GT
Max. speed
HiCen GT
Max. RCF
20 97 100 AS 2.6 2–6 microplates with two 80 mm-high carriers 4,000 rpm 1,970 ×g
20 90 501 AS 4.500 4×500 ml base rotor; buckets ordered separately — —
20 90 540 Carrier 4×microplates, set of four 4,000 rpm 2,720 ×g
20 90 502 Bucket 4×500 ml round buckets, set of four 4,500 rpm 3,960 ×g
20 90 550 Trunnion ring 4×(7×50 ml Falcon), set of four 4,500 rpm 4,090 ×g
20 90 551 Trunnion ring 4×(14×15 ml Falcon), set of four 4,500 rpm 4,070 ×g
20 70 060 AS 4.750 4×750 ml base rotor; buckets ordered separately — —
20 70 400 Carrier 4×microplates, set of four 3,700 rpm 2,690 ×g
20 70 300 Bucket 4×750 ml round buckets, set of four 3,000 rpm 2,130 ×g
20 70 050 AS 4.1000 4×1000 ml base rotor; buckets ordered separately — —
20 70 400 Carrier 4×microplates, set of four 3,700 rpm 2,690 ×g
20 70 300 Bucket 4×750 ml round buckets, set of four 2,500 rpm 1,480 ×g
20 70 200 Bucket 4×1000 ml round buckets, set of four 2,500 rpm 1,620 ×g
20 70 650 Trunnion ring 4×(8×50 ml Falcon), set of four 2,500 rpm 1,370 ×g
20 70 615 Trunnion ring 4×(19×15 ml Falcon), set of four 2,500 rpm 1,410 ×g

Angle rotors

Part number Rotor model Maximum capacity Angle HiCen GT
Max. speed
HiCen GT
Max. RCF
20 96 100 AF 24.2 24×1.5/2.2 ml 45° 14,000 rpm 21,910 ×g
20 75 100 AF 72.2 72×1.5/2.2 ml 40° 8,000 rpm 9,300 ×g
20 94 100 AF 20.10 20×10 ml 25° 11,000 rpm 13,260 ×g
20 26 100 AF 24.16 24×16 ml 28° 9,000 rpm 9,960 ×g
20 45 100 AF 12.30 12×30 ml 28° 12,000 rpm 17,070 ×g
20 93 005 AF 8.50.1 8×50 ml 34° 10,000 rpm 12,070 ×g
20 93 100 AF 8.50.2 8×50 ml 25° 9,000 rpm 8,690 ×g
20 98 105 AF 8.50.3 8×50 ml Falcon 40° 9,000 rpm 9,420 ×g
20 44 105 AF 12.50 12×50 ml 34° 7,000 rpm 7,010 ×g
20 92 105 AF 6.94 6×94 ml 25° 10,000 rpm 10,960 ×g
20 22 100 AF 6.250 6×250 ml 30° 7,000 rpm 7,940 ×g
20 23 100 AF 6.500 6×500 ml 30° 7,000 rpm 9,090 ×g
20 24 100 AF 4.1000 4×1000 ml 22° 4,500 rpm 3,850 ×g

Rotors, buckets, adaptors and tubes are selected as a system for the intended application; a dash indicates that the brochure does not list a standalone operating value for that component.

Key features

  • Up to 4×1000 ml with either swing-out or angle rotors.
  • Maximum speed of 14,000 rpm and 21,910×g.
  • Benchtop format for larger batches that do not require refrigeration.

Control and safety

  • Colour TFT touchscreen, APRI™ automatic rotor identification and storage for 50 programs.
  • Self-centring brushless drive with multistage acceleration, deceleration and dynamic electronic braking.
  • Imbalance shutdown, overspeed checking, automatic lid locking and emergency slow release.
  • Stainless-steel chamber for cleaning; refrigerated models support pre-cooling and continuous cooling.

Rotor and configuration selection

Red Line offers swing-out and angle-rotor configurations for bottles, Falcon tubes and multiple tube formats. Inert-gas protection is available for selected configurations; final rotors, adaptors and options are confirmed with the order.

Applications

For large-capacity clarification, cell harvesting, pelleting and production quality control at ambient temperature.