A.KRÜSS Optronic Tips correct cleaning refractometer
PDF · Manufacturer document
A refractometer prism is in direct contact with the sample. Films, dried deposits or cleaning liquid left by a previous measurement can change the conditions of the next measurement. A surface that looks clean is not necessarily ready for the next sample.
This article draws on the A.KRÜSS Optronic cleaning guide to explain prism cleaning and checks during sample changes. The cleaning medium, quantity and procedure must still follow the model instructions, sample properties and laboratory safety requirements.
The manufacturer recommends cleaning the prism immediately after each measurement to prevent samples drying or hardening on it. Delayed cleaning can make viscous residues harder to remove. Acidic, alkaline or otherwise aggressive samples particularly require prompt cleaning and a check of the prism condition.[1]
For sequential measurements, cleaning aims to remove visible contamination and reduce carryover into the next sample. Include cleaning checks when changing between aqueous and oily samples, or between samples with substantially different compositions.
The guide lists water for several aqueous sample types, including sugar solutions, salt solutions and honey, and suggests isopropanol for mineral oils. These examples show why wiping with water is not a universal cleaning method.[1]
The sample table is not a cross-model solvent recipe. Before using an organic solvent, check compatibility with the prism, seals and other wetted materials and consult its safety data sheet (SDS). Do not substitute a solvent merely because it seems stronger, or mix cleaning agents.
A reference check assesses the current measurement state; it does not mean the instrument must be readjusted each time. If a reference result is abnormal, first check the reference material, cleanliness and measurement conditions. Do not adjust the instrument while residues remain.
The manufacturer warns that bubbles affect results. Cleanliness and sample loading therefore need separate checks; not every fluctuation is a matter of instrument accuracy.[1]
Use a sequence of sample removal, appropriate cleaning, confirmation of cleanliness and smooth sample loading. A consistent cleaning and check procedure for the sample type makes troubleshooting easier than frequently changing solvents or repeatedly adjusting the instrument.