Herolab HiCen TR Inert Gas Purging Centrifuge System
Herolab

Herolab HiCen TR Inert Gas Purging Centrifuge System

Series: Inert Gas Purge System

Model: HiCen TR with N₂ purge

HiCen TR nitrogen-purged centrifuge system for oxidation-sensitive, flammable and other sensitive samples, with oxygen monitoring and safety interlocks.

Application / measured object:

HiCen TR nitrogen-purged centrifuge system for oxidation-sensitive, flammable and other sensitive samples, with oxygen monitoring and safety interlocks.

Application area:

Product overview

Built around the refrigerated HiCen TR platform, the system uses controlled nitrogen purging to reduce oxygen inside the rotor chamber and create a protected atmosphere for oxygen-sensitive or potentially flammable samples.

Technical specifications

Item Specification or description
Base centrifuge Refrigerated HiCen TR
Maximum speed Up to 15,000 rpm
Maximum RCF Up to 26,790 ×g
Maximum capacity 4×500 ml swing-out; 6×250 ml angle rotor
Rotor types Swing-out and fixed-angle rotors
Drive Maintenance-free brushless induction drive
Rotor chamber Reinforced stainless-steel chamber
Cooling Refrigerated
Safety equipment Nitrogen overpressure protection, oxygen monitoring, automatic emergency shutdown and motorised lid locking

Safety technology

  • Nitrogen overpressure protection and an oxygen sensor continuously monitor the rotor chamber.
  • Insufficient nitrogen, elevated oxygen concentration or incorrect connections prevent start-up or stop the run automatically.
  • Separate power supplies and series-connected safety interlocks increase protection reliability.
  • A bypass provides power-independent emergency nitrogen purging during a power failure.
  • Key safety components are configured to SIL 2 safety-integrity requirements.

Operating principle

Nitrogen is supplied before and during centrifugation to reduce oxygen inside the chamber. Pressure and oxygen monitoring jointly determine whether operation is enabled. Gas supply, exhaust, safety interlocks and site installation are confirmed for each project.

Applications

For sensitive-sample processing in pharmaceutical and biotechnology, chemical and materials research, industrial and environmental analysis, food and beverage, and life-science laboratories requiring a protective atmosphere.

Configuration note

This is a project-specific combination of the HiCen TR and the inert-gas protection system. Rotors, adaptors, nitrogen supply, exhaust connections and safety requirements are subject to final Labsun technical and order confirmation.